Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Goerz SE120: (1 year) CAL VER /5500 INSTRUMENT: Goerz SE124: (1 year) CAL VER /5500 DATE: 04-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 6 NUMBER OF LINES: 85 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # Sub Goerz SE120/SE124 (1 year) CAL VER /5500 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. 1.006 MEM2 =Channel 1 1.007 CALL Sub Goerz SE120/SE124 (1 year) CAL VER /5500 1.008 JMP 2.001 1.009 EVAL Dummy 2.001 MEM2 =Channel 2 2.002 CALL Sub Goerz SE120/SE124 (1 year) CAL VER /5500 2.003 JMP 3.001 2.004 EVAL Dummy 3.001 MEM2 =Channel 3 3.002 CALL Sub Goerz SE120/SE124 (1 year) CAL VER /5500 3.003 HEAD {DISTANCE BETWEEN RECORDING PENS} 3.004 DISP Use the ZERO knob on the measuring units to shift the 3.004 DISP trace to the right-hand & left-hand stops. Repeat this 3.004 DISP process on each channel. 3.005 MESS Refering to the above traces: 3.005 MESS 3.005 MESS The distance between recorded traces between channels 3.005 MESS must be 2.5mm +/-.2mm. 3.005 MESS 3.005 MESS The recording pens must not obstruct each other. 3.005 MESS 3.006 EVAL Are {the recording pens distances correct}? 4.001 HEAD {PEN SYNCH VERIFICATION} 4.002 DISP Connect the 5500A to the UUT as follows: 4.002 DISP [32] 5500A NORMAL HI to Channel 1, 2, and 3 INPUT HI 4.002 DISP [32] 5500A NORMAL LO to Channel 1, 2, and 3 INPUT LO 4.003 5500 0V S 2W 4.004 DISP Set the RANGE switch on all channels to 10V. 4.005 DISP Set all the pens to 50% using the ZERO control. 4.005 DISP Set the PAPER FEED to 60cm/min. 4.005 DISP Set the CHART FEED button to ON. 4.006 5500 4V 0.1H SI S 2W 4.007 MESS With the PEN SYNCH OFF verify that the channel traces 4.007 MESS are separated on the chart by approx. 2.5mm. 4.007 MESS 4.008 EVAL Are {the traces separated by approx 2.5mm w/PEN SYNCH OFF}? 5.001 DISP Set the PEN SYNCH to ON. 5.002 MESS With the pen synch ON, channels 2&3 are stored 5.002 MESS temporarily and output with a time delay. The delay 5.002 MESS depends on the chart speed and channel number. 5.002 MESS 5.003 EVAL Are {the traces on top of each other w/PEN SYNCH ON}? 6.001 DISP Set the CHART FEED button to OFF. 6.002 END